Back to Search
Start Over
Continuous Glucose Monitoring in the Intensive Care Unit Following Total Pancreatectomy with Islet Autotransplantation in Children: Establishing Accuracy of the Dexcom G6 Model.
- Source :
-
Journal of clinical medicine [J Clin Med] 2021 Apr 27; Vol. 10 (9). Date of Electronic Publication: 2021 Apr 27. - Publication Year :
- 2021
-
Abstract
- Hyperglycemia is detrimental to postoperative islet cell survival in patients undergoing total pancreatectomy with islet autotransplantation (TPIAT). This makes continuous glucose monitoring (CGM) a useful management tool. We evaluated the accuracy of the Dexcom G6 CGM in pediatric intensive care unit patients following TPIAT. Twenty-five patients who underwent TPIAT had Dexcom G6 glucose values compared to paired serum glucose values. All paired glucose samples were obtained within 5 minutes of each other during the first seven days post TPIAT. Data were evaluated using mean absolute difference (MAD), mean absolute relative difference (MARD), %20/20, %15/15 accuracy, and Clarke Error Grid analysis. Exclusions included analysis during the CGM "warm-up" period and hydroxyurea administration (known drug interference). A total of 183 time-matched samples were reviewed during postoperative days 2-7. MAD was 14.7 mg/dL and MARD was 13.4%, with values of 15.2%, 14.0%, 12.1%, 11.4%, 13.2% and 14.1% at days 2, 3, 4, 5, 6 and 7, respectively. Dexcom G6 had a %20/20 accuracy of 78%, and a %15/15 accuracy of 64%. Clarke Error Grid analysis showed that 77% of time-matched values were clinically accurate, and 100% were clinically acceptable. The Dexcom G6 CGM may be an accurate tool producing clinically acceptable values to make reliable clinical decisions in the immediate post-TPIAT period.
Details
- Language :
- English
- ISSN :
- 2077-0383
- Volume :
- 10
- Issue :
- 9
- Database :
- MEDLINE
- Journal :
- Journal of clinical medicine
- Publication Type :
- Academic Journal
- Accession number :
- 33925523
- Full Text :
- https://doi.org/10.3390/jcm10091893