Calibration factors for Ge detectors used for field spectrometry.

Authors :
Helfer IK
Miller KM
Source :
Health physics [Health Phys] 1988 Jul; Vol. 55 (1), pp. 15-29.
Publication Year :
1988

Abstract

Calibration factors to convert a measured full-absorption peak count rate to activity in the soil and dose rate in air are given for Ge detectors that are used for field measurements of radionuclides. The appropriate factors for a given detector are derived using three primary parameters: the manufacturer's quoted efficiency at 1332 keV relative to a 7.6 cm (3 in.) long by 7.6 cm (3 in.) diameter NaI(Tl) detector, the detector's orientation in the field (up or down) and the Ge crystal length/diameter ratio. The accuracy of the results obtained by using this simplified calibration technique is estimated to be 10-15%.