Magnetocardiograms (MCG) provide clinically useful diagnostic information in a variety of cardiac dysfunctions. Low frequency baseline drifts and high frequency noise are inevitably present in routine MCG even for those measured inside magnetically shielded rooms. These interferences sometimes exceed subtle cardiac features in MCG recorded on subjects with implanted devices like cardiac pacemakers; this makes interpretation of cardiac magnetic fields difficult. The present study proposes a correlation-based beat-by-beat approach and principal component analysis to eliminate drifts and high frequency noise respectively; the approach is suitable for denoising both single and multi-channel MCG data. The methodology is critically evaluated on simulated noisy measurements using a 37 channel MCG system, when objects such as implantable permanent pacemaker and stainless-steel wire are sequentially kept externally on the chests of five healthy subjects. By characterizing the noise introduced by each of these objects, the deterioration in the quality of MCG and its subsequent restoration by using the proposed method is assessed. The performance of the proposed method is also compared with other conventional denoising techniques namely, bandpass filters, wavelets and ensemble empirical mode decomposition. The proposed method not only exhibits least distortion, but also preserves the beat-by-beat dynamics of cardiac time series. The method has also been illustrated on actual MCG measurements on two subjects with implanted pacemaker which highlight the ability of the proposed method for denoising MCG in general and during extremely noisy measurement situations.<br /> (© 2021 IOP Publishing Ltd.)

Details

Language :
English
ISSN :
2057-1976
Volume :
7
Issue :
3
Database :
MEDLINE
Journal :
Biomedical physics & engineering express
Publication Type :
Academic Journal
Accession number :
33662938
Full Text :
https://doi.org/10.1088/2057-1976/abec17