Back to Search Start Over

Validation of an artificial intelligence-driven large vessel occlusion detection algorithm for acute ischemic stroke patients.

Authors :
Rava RA
Peterson BA
Seymour SE
Snyder KV
Mokin M
Waqas M
Hoi Y
Davies JM
Levy EI
Siddiqui AH
Ionita CN
Source :
The neuroradiology journal [Neuroradiol J] 2021 Oct; Vol. 34 (5), pp. 408-417. Date of Electronic Publication: 2021 Mar 03.
Publication Year :
2021

Abstract

Rapid and accurate diagnosis of large vessel occlusions (LVOs) in acute ischemic stroke (AIS) patients using automated software could improve clinical workflow in determining thrombectomy in eligible patients. Artificial intelligence-based methods could accomplish this; however, their performance in various clinical scenarios, relative to clinical experts, must be thoroughly investigated. We aimed to assess the ability of Canon's <superscript>AUTO</superscript> Stroke Solution LVO application in properly detecting and locating LVOs in AIS patients. Data from 202 LVO and 101 non-LVO AIS patients who presented with stroke-like symptoms between March 2019 and February 2020 were collected retrospectively. LVO patients had either an internal carotid artery (ICA) ( n  = 59), M1 middle cerebral artery (MCA) ( n  = 82) or M2 MCA ( n  = 61) occlusion. Computed tomography angiography (CTA) scans from each patient were pushed to the automation platform and analyzed. The algorithm's ability to detect LVOs was assessed using accuracy, sensitivity and Matthews correlation coefficients (MCCs) for each occlusion type. The following results were calculated for each occlusion type in the study (accuracy, sensitivity, MCC): ICA = (0.95, 0.90, 0.89), M1 MCA = (0.89, 0.77, 0.78) and M2 MCA = (0.80, 0.51, 0.59). For the non-LVO cohort, 98% (99/101) of cases were correctly predicted as LVO negative. Processing time for each case was 69.8 ± 1.1 seconds (95% confidence interval). Canon's <superscript>AUTO</superscript> Stroke Solution LVO application was able to accurately identify ICA and M1 MCA occlusions in addition to almost perfectly assessing when an LVO was not present. M2 MCA occlusion detection needs further improvement based on the sensitivity results displayed by the LVO detection algorithm.

Details

Language :
English
ISSN :
2385-1996
Volume :
34
Issue :
5
Database :
MEDLINE
Journal :
The neuroradiology journal
Publication Type :
Academic Journal
Accession number :
33657922
Full Text :
https://doi.org/10.1177/1971400921998952