Back to Search Start Over

Mapping Cell Viability Quantitatively and Independently From Cell Density in 3D Gels Noninvasively.

Authors :
Archer BJ
Mack JJ
Acosta S
Nakasone R
Dahoud F
Youssef K
Goldstein A
Goldsman A
Held MC
Wiese M
Blumich B
Wessling M
Emondts M
Klankermayer J
Iruela-Arispe ML
Bouchard LS
Source :
IEEE transactions on bio-medical engineering [IEEE Trans Biomed Eng] 2021 Oct; Vol. 68 (10), pp. 2940-2947. Date of Electronic Publication: 2021 Sep 20.
Publication Year :
2021

Abstract

Objective: In biomanufacturing there is a need for quantitative methods to map cell viability and density inside 3D bioreactors to assess health and proliferation over time. Recently, noninvasive MRI readouts of cell density have been achieved. However, the ratio of live to dead cells was not varied. Herein we present an approach for measuring the viability of cells embedded in a hydrogel independently from cell density to map cell number and health.<br />Methods: Independent quantification of cell viability and density was achieved by calibrating the <superscript>1</superscript> H magnetization transfer- (MT) and diffusion-weighted NMR signals to samples of known cell density and viability using a multivariate approach. Maps of cell viability and density were generated by weighting NMR images by these parameters post-calibration.<br />Results: Using this method, the limits of detection (LODs) of total cell density and viable cell density were found to be 3.88 ×10 <superscript>8</superscript> cells · mL <superscript>-1</superscript> · Hz <superscript>-1/2</superscript> and 2.36 ×10 <superscript>9</superscript> viable cells · mL <superscript>-1</superscript> · Hz <superscript>-1/2</superscript> respectively.<br />Conclusion: This mapping technique provides a noninvasive means of visualizing cell viability and number density within optically opaque bioreactors.<br />Significance: We anticipate that such nondestructive readouts will provide valuable feedback for monitoring and controlling cell populations in bioreactors.

Details

Language :
English
ISSN :
1558-2531
Volume :
68
Issue :
10
Database :
MEDLINE
Journal :
IEEE transactions on bio-medical engineering
Publication Type :
Academic Journal
Accession number :
33531296
Full Text :
https://doi.org/10.1109/TBME.2021.3056526