Development of capillary-paper spray for small-molecule analysis in complex samples.

Authors :
Liu C
Li Y
Xiong C
Nie Z
Source :
Analytical and bioanalytical chemistry [Anal Bioanal Chem] 2021 Feb; Vol. 413 (4), pp. 1099-1106. Date of Electronic Publication: 2021 Jan 03.
Publication Year :
2021

Abstract

We develop a capillary-paper spray (CPS) ion source which allows for sample separation in the capillary and enables rapid and sensitive paper spray (PS) mass spectrometry (MS) analysis of biofluids. The CPS employs a glass capillary to load liquid analytes, vertically standing at the rear of the PS. To further reduce the matrix effect, a nitrocellulose filter membrane is placed between the glass tube and chromatography paper to absorb proteins and other macromolecules, which is beneficial for the detection of the small molecules. Compared with the normal PS method, the CPS method markedly improves spray stability and prolongs analysis duration, and also generates significantly better signal intensities during the analysis of drugs, thus indicating its potential for clinical use. As a proof of concept, quantitative analysis of drugs (metformin hydrochloride and berberine hydrochloride) in serum is performed.