Back to Search Start Over

Analysis and suppression of the polarization error for the optical rotation detection system in an atomic comagnetometer.

Authors :
Huang J
Wang Z
Fan W
Xing L
Zhang W
Duan L
Quan W
Source :
Optics express [Opt Express] 2020 Nov 23; Vol. 28 (24), pp. 35748-35760.
Publication Year :
2020

Abstract

This paper investigates the laser polarization error in the optical rotation detection system (ORDS) of an atomic comagnetometer (ACM), which will seriously degrade the long-term performance of the ORDS. We first establish an optical transmission model of the ORDS by using Jones matrix concerning the optical imperfection of polarizers. Then, we analyze the polarization error based on this model and propose a novel error suppression method. Finally, we experimentally test the long-term performance of the ORDS and the ACM before and after the polarization error suppression to verify the effectiveness of the proposed method. The experimental results show that the long-term performance of the ORDS and the ACM can be improved by approximately 3.4 times with the proposed polarization error suppression method.

Details

Language :
English
ISSN :
1094-4087
Volume :
28
Issue :
24
Database :
MEDLINE
Journal :
Optics express
Publication Type :
Academic Journal
Accession number :
33379685
Full Text :
https://doi.org/10.1364/OE.406073