An Improved Reflection Colorimeter Integrated with a Coaxial Optical-fiber Cable for Highly Sensitive Solid-phase Colorimetry Using a Membrane Filter.

Authors :
Mizuguchi H
Taniguchi Y
Suherman S
Morita K
Takayanagi T
Suzuki Y
Source :
Analytical sciences : the international journal of the Japan Society for Analytical Chemistry [Anal Sci] 2021 Jul 10; Vol. 37 (7), pp. 1045-1048. Date of Electronic Publication: 2020 Dec 11.
Publication Year :
2021

Abstract

Highly sensitive solid-phase colorimetry for nickel ion was demonstrated using an improved reflection colorimeter equipped with a coaxial optical-fiber cable. The nickel complex with α-furil dioxime was collected on a small-size membrane filter embedded in a disposable syringe filter unit. The leading edge of the optical-fiber cable was connected to the syringe filter unit via a Luer-lock fitting, and the color intensity of the sample on the filter was evaluated accurately. The detection limit was 0.8 ng in 2.5 mL of the complex solution (0.3 μg L <superscript>-1</superscript> ). This improved configuration is applicable to highly sensitive on-site analysis without expensive instruments nor high laboratory skills.