Back to Search Start Over

Pre-analytical and analytical variables that influence urinary volatile organic compound measurements.

Authors :
McFarlanE M
MozdiaK E
Daulton E
Arasaradnam R
Covington J
Nwokolo C
Source :
PloS one [PLoS One] 2020 Jul 31; Vol. 15 (7), pp. e0236591. Date of Electronic Publication: 2020 Jul 31 (Print Publication: 2020).
Publication Year :
2020

Abstract

There has been rapidly accelerating interest in the utilization of volatile organic compounds (VOCs) as non-invasive methods for rapid point-of-care medical diagnostics. There is widespread variation in analytical methods and protocols, with little understanding of the effects of sample storage on VOC profiles. This study aimed to determine the effects on VOC profiles of different storage times, at room temperature, prior to freezing, of sealed urine samples from healthy individuals. Analysis using Field Asymmetric Ion Motility Spectrometry (FAIMS) determined the alterations in VOC and total ion count profiles as a result of increasing room temperature storage times. Results indicated that increasing exposure time to room temperature prior to freezing had a threefold effect. Firstly, increased urinary VOC profile variability, with a plateau phase between 12 and 48 hours, before further degradation. Secondly, an increase in total ion count with time exposed to room temperature. Finally, a deterioration in VOCs with each sample run during the analysis process. This provides new insight into the effect of storage of urine samples for VOC analysis using FAIMS technology. Results of this study provide a recommendation for a 12-hour maximum duration at room temperature prior to storage.<br />Competing Interests: The authors have declared that no competing interests exist.

Details

Language :
English
ISSN :
1932-6203
Volume :
15
Issue :
7
Database :
MEDLINE
Journal :
PloS one
Publication Type :
Academic Journal
Accession number :
32735600
Full Text :
https://doi.org/10.1371/journal.pone.0236591