Back to Search
Start Over
3D-printed bolus improves dose distribution for veterinary patients treated with photon beam radiation therapy.
- Source :
-
The Canadian veterinary journal = La revue veterinaire canadienne [Can Vet J] 2020 Jun; Vol. 61 (6), pp. 638-644. - Publication Year :
- 2020
-
Abstract
- Commercial bolus is frequently used to increase dose at the patient's surface for superficial radiotherapy; however, uneven surfaces can create air gaps and discrepancies between prescribed and delivered dose. The purpose of this study was to determine if a customizable, 3D-printed bolus would improve dosimetry compared with a commercial bolus. For each patient, a planned bolus was generated within planning software, then created with 3D-printing. The treatment plan was recalculated with each bolus in situ . When evaluating tumor volumes at prescription, the 3D-printed bolus was closer to prescription compared to the commercial bolus. There was a significant difference in air gaps in patients receiving radiotherapy to the head ( P < 0.001) but the difference was not significant for air gaps in caudal body sites ( P = 0.05). Overall, the 3D-printed bolus resulted in reduced air gaps, dosimetry closer to prescription, and should be considered for superficial treatment areas of high irregularity.<br /> (Copyright and/or publishing rights held by the Canadian Veterinary Medical Association.)
Details
- Language :
- English
- ISSN :
- 0008-5286
- Volume :
- 61
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- The Canadian veterinary journal = La revue veterinaire canadienne
- Publication Type :
- Academic Journal
- Accession number :
- 32675816