Back to Search
Start Over
Duplex Ultrasound Investigation for the Detection of Obstructed Iliocaval Venous Stents.
- Source :
-
European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery [Eur J Vasc Endovasc Surg] 2020 Sep; Vol. 60 (3), pp. 443-450. Date of Electronic Publication: 2020 Jun 27. - Publication Year :
- 2020
-
Abstract
- Objective: Duplex ultrasound (DUS) is used for routine surveillance of stents in iliocaval veins, but direct visualisation is often challenging. Duplex ultrasound criteria for detecting venous stent obstruction (VSO) have not been defined to date.<br />Methods: A nested case control study of 120 patients (42 ± 17 years, 53% women, mean 2.7 ± 1.8 stents) was performed, and the performance of various duplex parameters for detecting VSO (defined as > 50% lumen diameter reduction or occlusion) was tested, confirmed by biplane venography or intravascular ultrasound (IVUS). Forty patients with VSO (25 with stent occlusion, 15 with >50% in stent stenosis) were matched to 80 control patients by age, gender and index diagnosis who fulfilled the following criteria: (1) ongoing symptom control (Villalta score < 5), (2) good image quality of entire stent segment, (3) spontaneous colour Doppler signal > 50% of lumen in entire stent segment, (4) at least two DUS where the baseline DUS was obtained within 24 h after successful venous intervention.<br />Results: The best test was the combination of peak flow velocity and flow pattern analysis at the stent inlet. A peak flow velocity >10 cm/s and a flow pattern spontaneously modulated by respiration ruled out VSO with a specificity of 93.7% (95% CI 86.0%-97.3%). A peak flow velocity ≤10 cm/s or any Doppler flow pattern other than spontaneously modulated by respiration was 92.1% (95% CI 79.2%-97.3%) sensitive to detect VSO.<br />Conclusion: The combination of peak flow velocity and analysis of Doppler flow pattern at the stent inlet is accurate to diagnose or rule out stent occlusion. Indirect criteria should always be combined with direct visualisation of iliocaval stents since those may be less sensitive for detecting stent stenosis.<br /> (Copyright © 2020 European Society for Vascular Surgery. Published by Elsevier B.V. All rights reserved.)
- Subjects :
- Adult
Blood Flow Velocity
Databases, Factual
Endovascular Procedures adverse effects
Female
Humans
Iliac Vein physiopathology
Male
Middle Aged
Phlebography
Predictive Value of Tests
Registries
Retrospective Studies
Treatment Outcome
Vascular Diseases diagnostic imaging
Vascular Diseases physiopathology
Vascular Patency
Vena Cava, Inferior physiopathology
Young Adult
Endovascular Procedures instrumentation
Iliac Vein diagnostic imaging
Stents
Ultrasonography, Doppler, Color
Vascular Diseases therapy
Vena Cava, Inferior diagnostic imaging
Subjects
Details
- Language :
- English
- ISSN :
- 1532-2165
- Volume :
- 60
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery
- Publication Type :
- Academic Journal
- Accession number :
- 32605854
- Full Text :
- https://doi.org/10.1016/j.ejvs.2020.05.011