SAPIEN S3 valve deployment in the pulmonary position using the gore DrySeal sheath to protect the tricuspid valve.
- Source :
-
Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions [Catheter Cardiovasc Interv] 2020 Nov; Vol. 96 (6), pp. 1287-1293. Date of Electronic Publication: 2020 Jun 30. - Publication Year :
- 2020
-
Abstract
- Background: Tricuspid valve injury can occur during implantation of a SAPIEN valve in the pulmonary position. We describe our experience using a long Gore DrySeal (GDS) sheath to protect the tricuspid valve during advancement of the Commander delivery system.<br />Methods: Retrospective single center review of all patients who underwent placement of a SAPIEN valve in the right ventricular outflow tract between January 2016 and April 2020. Patients were divided into two groups: delivery of the valve using standard technique (Group I), and with the use of a GDS (Group II), for comparison.<br />Results: There were 48 patients in total: 25 in Group I and 23 in Group II. In Group II, the first 10 patients had a 29 mm S3 placed through a 26 French (Fr), 65 cm GDS. We then performed additional crimping of the S3 onto the balloon after the balloon catheter was withdrawn to position the valve on the balloon outside the body. Subsequently, seven had a 29 mm S3 placed through a 24 Fr GDS, and four had a 26 mm S3 placed through a 22 Fr GDS including one weighing 16 kg. Two had a 23 mm S3 placed through a 22Fr GDS as the 20Fr GDS was not available in our lab. Severe tricuspid valve injury occurred in 2/25 (8%) of Group I patients and 0/23 of Group II patients.<br />Conclusion: Use of a long GDS may protect the tricuspid valve from injury during implantation of the S3 valve in the pulmonary position, and is technically feasible in smaller patients.<br /> (© 2020 Wiley Periodicals LLC.)
- Subjects :
- Adolescent
Adult
Balloon Valvuloplasty
Cardiac Catheterization adverse effects
Child
Female
Heart Injuries diagnostic imaging
Heart Injuries etiology
Heart Injuries physiopathology
Heart Valve Prosthesis Implantation adverse effects
Humans
Male
Prosthesis Design
Pulmonary Valve diagnostic imaging
Pulmonary Valve physiopathology
Retrospective Studies
Risk Assessment
Risk Factors
Time Factors
Treatment Outcome
Tricuspid Valve diagnostic imaging
Tricuspid Valve injuries
Tricuspid Valve Insufficiency diagnostic imaging
Tricuspid Valve Insufficiency etiology
Tricuspid Valve Insufficiency physiopathology
Young Adult
Cardiac Catheterization instrumentation
Heart Injuries prevention & control
Heart Valve Prosthesis
Heart Valve Prosthesis Implantation instrumentation
Hemodynamics
Pulmonary Valve surgery
Tricuspid Valve physiopathology
Tricuspid Valve Insufficiency prevention & control
Subjects
Details
- Language :
- English
- ISSN :
- 1522-726X
- Volume :
- 96
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
- Publication Type :
- Academic Journal
- Accession number :
- 32602984
- Full Text :
- https://doi.org/10.1002/ccd.29120