Back to Search
Start Over
Wide-field polygon-scanning photoacoustic microscopy of oxygen saturation at 1-MHz A-line rate.
- Source :
-
Photoacoustics [Photoacoustics] 2020 Jun 07; Vol. 20, pp. 100195. Date of Electronic Publication: 2020 Jun 07 (Print Publication: 2020). - Publication Year :
- 2020
-
Abstract
- We report wide-field polygon-scanning functional OR-PAM that for the first time achieves 1-MHz A-line rate of oxygen saturation in vivo. We address two technical challenges. The first is a 1-MHz dual-wavelength pulsed laser that has sufficient pulse energy and ultrafast wavelength switching. The second is a polygon-scanning imaging probe that has a fast scanning speed, a large field of view, and great sensitivity. The OR-PAM system offers a B-scan rate of 477.5 Hz in a 12-mm range and a volumetric imaging rate of ∼1 Hz over a 12 × 5 mm <superscript>2</superscript> scanning area. We image microvasculature and blood oxygen saturation in a 12 × 12 mm <superscript>2</superscript> scanning area in 5 s. Dynamic imaging of oxygen saturation in the mouse ear is demonstrated to monitor fast response to epinephrine injection. The new wide-field fast functional imaging ability broadens the biomedical application of OR-PAM.<br />Competing Interests: The authors declare no conflicts of interest.<br /> (© 2020 The Authors.)
Details
- Language :
- English
- ISSN :
- 2213-5979
- Volume :
- 20
- Database :
- MEDLINE
- Journal :
- Photoacoustics
- Publication Type :
- Academic Journal
- Accession number :
- 32577378
- Full Text :
- https://doi.org/10.1016/j.pacs.2020.100195