Dosimetry formalism and calibration procedure for electronic brachytherapy sources in terms of absorbed dose to water.
- Source :
-
Physics in medicine and biology [Phys Med Biol] 2020 Jul 27; Vol. 65 (14), pp. 145006. Date of Electronic Publication: 2020 Jul 27. - Publication Year :
- 2020
-
Abstract
- The LNE-LNHB has developed a methodology to standardize electronic brachytherapy sources in terms of absorbed dose to water. It is based on the measurement of the air-kerma rate at a given distance from the source and the Monte Carlo calculation of a conversion factor. This factor converts the air-kerma in measurement conditions into absorbed dose to water at a 1 cm reference depth in a water phantom. As a first application, the method was used to calibrate a Zeiss INTRABEAM system equipped with its 4 cm diameter spherical applicator. The absorbed-dose rate value obtained in the current study was found significantly higher than that provided by the manufacturer in line with the observations already reported by a few other teams.