Leukoreduction in ex vivo perfusion circuits: comparison of leukocyte depletion efficiency with leukocyte filters.

Authors :
Fahradyan V
Annunziata MJ
Said S
Rao M
Shah H
Ordenana C
Papay FA
Rampazzo A
Bassiri Gharb B
Source :
Perfusion [Perfusion] 2020 Nov; Vol. 35 (8), pp. 853-860. Date of Electronic Publication: 2020 May 13.
Publication Year :
2020

Abstract

Background: Leukodepletion of whole blood-based perfusates remains a challenge in experimental models of ex vivo perfusion. This study investigated the leukoreduction efficacy of the commonly used LeukoGuard LG Arterial and BC2 Cardioplegia filters.<br />Methods: Eleven liters of washed porcine blood was used to evaluate the filtration efficiency of LG (n = 6) and BC2 (n = 5) filters. Filter efficacy was tested by passing 1 L of washed blood through each filter. Complete blood count was performed to detect a reduction of white blood cells, red blood cells, and hemoglobin concentration.<br />Results: The BC2 Cardioplegia filter showed a significant reduction in white blood cell count (13.16 ± 4.2 × 10 <superscript>3</superscript>  cells/μL pre-filtration, 0.62 ± 0.61 cells/μL post-filtration, p = 0.005), red blood cell count (9.18 ± 0.16 × 10 <superscript>6</superscript>  cells/μL pre-filtration, 9.02 ± 0.16 × 10 <superscript>6</superscript>  cells/μL post-filtration, p = 0.012) and hemoglobin concentration (15.89 ± 0.66 g/dL pre-filtration, 15.67 ± 0.83 g/dL post-filtration, p = 0.017). Platelet reduction in the LG filter group was statistically significant (13.23 ± 13.98 × 10 <superscript>3</superscript>  cells/μL pre-filtration, 7.15 ± 3.31 × 10 <superscript>3</superscript>  cells/μL post-filtration, p = 0.029), but no difference was seen in the BC2 group. There was no significant difference in white blood cell count in the LG filter group (10.12 ± 3.0 × 10 <superscript>3</superscript>  cells/μL pre-filtration, 10.32 ± 2.44 × 10 <superscript>3</superscript>  cells/μL post-filtration, p = 0.861).<br />Conclusion: Our results suggest that the LG filter should not be used in ex vivo perfusion circuits for the purpose of leukodepletion. The BC2 filter can be used in EVP circuits with flow rates of less than 350 mL/min. Alternatively, perfusate may be leukodepleted before perfusion.

Details

Language :
English
ISSN :
1477-111X
Volume :
35
Issue :
8
Database :
MEDLINE
Journal :
Perfusion
Publication Type :
Academic Journal
Accession number :
32404024
Full Text :
https://doi.org/10.1177/0267659120917872