Repeatability and Reproducibility of Photoreceptor Density Measurement in the Macula Using the Spectralis High Magnification Module.

Authors :
Mendonça LSM
Braun PX
Martin SM
Hüther A
Mehta N
Zhao Y
Abu-Qamar O
Konstantinou EK
Regatieri CVS
Witkin AJ
Baumal CR
Duker JS
Waheed NK
Source :
Ophthalmology. Retina [Ophthalmol Retina] 2020 Nov; Vol. 4 (11), pp. 1083-1092. Date of Electronic Publication: 2020 May 01.
Publication Year :
2020

Abstract

Purpose: To evaluate the repeatability and reproducibility of photoreceptor density assessment with manual cell counting in healthy participants imaged with the Heidelberg Spectralis High Magnification Module (HMM).<br />Design: Precision study, evaluation of diagnostic test or technology.<br />Participants: Eleven eyes of 8 participants.<br />Methods: Images were acquired using the Spectralis HMM by a single operator during 2 separate imaging sessions. The 3 highest-quality images of each eye from each session were selected for analysis and coregistered. For a subset of participants, a second operator acquired images in 1 session, and images with the best quality were selected for analysis. Photoreceptor densities were obtained by manual counts in squares of 0.0625 mm <superscript>2</superscript> located in the parafovea. Repeatability (intragrader and intrasession) and reproducibility (interoperator, intergrader, and intersession) were assessed by calculating the intraclass correlation coefficient (ICC) from linear mixed effects models. Bland-Altman plots, coefficients of repeatability, and Pearson correlation results were reported.<br />Main Outcome Measures: Intragrader, intrasession, intersession, interoperator, and intergrader ICC estimates and their 95% confidence intervals for photoreceptor density measurements in the parafovea.<br />Results: Twenty-four eyes of 13 healthy participants were imaged initially. Of these, 11 eyes (45.83%) of 8 participants that had at least 3 acceptable images in each session were included in this study. Mean parafoveal photoreceptor density was 14 988 cells/mm <superscript>2</superscript> (standard deviation, 1403.15 cells/mm <superscript>2</superscript> ). Intragrader ICC was 0.84 (95% confidence interval, 0.57-0.95), intrasession ICC was 0.69 (95% confidence interval, 0.17-0.86), intersession ICC was 0.88 (95% confidence interval, 0.53-0.96), interoperator ICC was 0.70 (95% confidence interval, 0-0.95), and intergrader ICC was 0.22 (95% confidence interval, 0-0.71).<br />Conclusions: Images obtained with the HMM allow for photoreceptor mosaic visualization in the macular area, mainly in the parafovea. Although densities obtained are in accordance with other reported methods in the literature, variability within and between images of the apparent cell mosaic were observed, and this study did not demonstrate high repeatability or reproducibility for quantitative assessments using the manual counting method.<br /> (Copyright © 2020 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
2468-6530
Volume :
4
Issue :
11
Database :
MEDLINE
Journal :
Ophthalmology. Retina
Publication Type :
Academic Journal
Accession number :
32371125
Full Text :
https://doi.org/10.1016/j.oret.2020.04.021