Back to Search Start Over

A Sensitive XRF Screening Method for Lead in Drinking Water.

Authors :
Tighe M
Bielski M
Wilson M
Ruscio-Atkinson G
Peaslee GF
Lieberman M
Source :
Analytical chemistry [Anal Chem] 2020 Apr 07; Vol. 92 (7), pp. 4949-4953. Date of Electronic Publication: 2020 Mar 09.
Publication Year :
2020

Abstract

A novel method for quickly and quantitatively measuring aqueous lead in drinking water has been developed. A commercially available activated carbon felt has been found to effectively capture lead from tap water, and partnered with X-ray fluorescence (XRF) spectrometry, it provides quantitative measurement of aqueous lead in drinking water. Specifically, for a 2 L volume of tap water, the linear range of detection was found to be from 1-150 ppb, encompassing the current EPA limit for lead in drinking water (15 ppb). To make a reproducible and easy to use method for filtering, a 2 L bottle cap with a 1.25 cm diameter hole was used for filtering. Utilizing this filtration method, 75 solutions from 0 to 150 ppb lead gave a 91% sensitivity, 97% specificity, and 93% accuracy, and all the misclassified samples fell between 10 and 15 ppb. This method has also proved reliable for detecting calcium as well as several other divalent metals in drinking water including copper, zinc, iron, and manganese.

Details

Language :
English
ISSN :
1520-6882
Volume :
92
Issue :
7
Database :
MEDLINE
Journal :
Analytical chemistry
Publication Type :
Academic Journal
Accession number :
32100538
Full Text :
https://doi.org/10.1021/acs.analchem.9b05058