Back to Search Start Over

A photographic technique for quick assessment of mechanical isocenter of a linear accelerator.

Authors :
Heikkilä J
Seppälä J
Nuutinen J
Vanne A
Lahtinen T
Source :
Technical innovations & patient support in radiation oncology [Tech Innov Patient Support Radiat Oncol] 2017 Apr 28; Vol. 2, pp. 1-4. Date of Electronic Publication: 2017 Apr 28 (Print Publication: 2017).
Publication Year :
2017

Abstract

Demands for mechanical accuracy of medical linear accelerators are increased due to the stereotactic and modulated rotational treatments. Mechanical inaccuracies affect the size and shape of the mechanical and radiation isocenters. In practice, the mechanical isocenter is defined by the intersection of rotational axes. However, there are no simple tools to check the properties of the mechanical isocenter in 3D. We introduce a new photography-based method for quick and sub-millimeter accurate determination of the mechanical isocenter. The method is based on image-processing algorithm and modified front pointer. The results demonstrate the quick measurement and visualization of the mechanical isocenter.<br /> (© 2017 The Authors.)

Details

Language :
English
ISSN :
2405-6324
Volume :
2
Database :
MEDLINE
Journal :
Technical innovations & patient support in radiation oncology
Publication Type :
Academic Journal
Accession number :
32095557
Full Text :
https://doi.org/10.1016/j.tipsro.2017.03.001