Back to Search Start Over

Fully waterproof two-axis galvanometer scanner for enhanced wide-field optical-resolution photoacoustic microscopy.

Authors :
Lee J
Han S
Seong D
Lee J
Park S
Eranga Wijesinghe R
Jeon M
Kim J
Source :
Optics letters [Opt Lett] 2020 Feb 15; Vol. 45 (4), pp. 865-868.
Publication Year :
2020

Abstract

A large field-of-view and fast scanning of photoacoustic microscopy (PAM) relatively have been difficult to obtain due to the water-drowned structure of the system for the transmission of ultrasonic signals. Researchers have widely studied the achievement of a waterproof scanner for dynamic biological applications with a high-resolution and high signal-to-noise ratio. This Letter reports a novel, to the best of our knowledge, waterproof galvanometer scanner-based PAM system with a successfully attainable ${9.0}\;{\rm mm} \times {14.5}\;{\rm mm}$9.0mm×14.5mm scan region, amplitude scan rate of 40 kHz, and spatial resolution of 4.9 µm. The in vivo characterization of a mouse brain in intact-skull microvascular visualization demonstrated its capability in biomedical imaging and is anticipated to be an effective technique for various preclinical and clinical studies.

Details

Language :
English
ISSN :
1539-4794
Volume :
45
Issue :
4
Database :
MEDLINE
Journal :
Optics letters
Publication Type :
Academic Journal
Accession number :
32058491
Full Text :
https://doi.org/10.1364/OL.380032