Back to Search Start Over

Emerging electromagnetic interferences between implantable cardioverter-defibrillators and left ventricular assist devices.

Authors :
Yalcin YC
Kooij C
Theuns DAMJ
Constantinescu AA
Brugts JJ
Manintveld OC
Yap SC
Szili-Torok T
Bogers AJJC
Caliskan K
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2020 Apr 01; Vol. 22 (4), pp. 584-587.
Publication Year :
2020

Abstract

Aims: To investigate the prevalence of electromagnetic interference (EMI) between left ventricular assist devices (LVADs) and implantable cardioverter-defibrillators (ICDs)/pacemakers (PMs).<br />Methods and Results: A retrospective single-centre study was conducted, including all patients undergoing HeartMate II (HMII) and HeartMate 3 (HM3) LVAD implantation (n = 106). Electromagnetic interference was determined by the inability to interrogate the ICD/PM. Overall, 85 (mean age 59 ± 8, 79% male) patients had an ICD/PM at the time of LVAD implantation; 46 patients with HMII and 40 patients with HM3. Among the 85 LVAD patients with an ICD's/PM's, 11 patients (13%) experienced EMI; 6 patients (15%) with an HMII and 5 patients (11%) with an HM3 (P = 0.59). Electromagnetic interference from the HMII LVADs was only present in patients with a St Jude/Abbott device; 6 of the 23 St Jude/Abbott devices. However, in the HM3 patients, EMI was mainly present in patients with Biotronik devices: 4 of the 18 with only one (1/25) patient with a Medtronic device. While initial interrogation of these devices was not successful, none of the 11 cases experienced pacing inhibition or inappropriate shocks.<br />Conclusion: In summary, the prevalence of EMI between ICDs in the older and newer type of LVAD's remains rather high. While HMII patients experienced EMI with a St Jude/Abbott device (which was already known), HM3 LVAD patients experience EMI mainly with Biotronik devices. Prospective follow-up, preferably in large registries, is warranted to investigate the overall prevalence and impact of EMI in LVAD patients.<br /> (© The Author(s) 2020. Published by Oxford University Press on behalf of the European Society of Cardiology.)

Details

Language :
English
ISSN :
1532-2092
Volume :
22
Issue :
4
Database :
MEDLINE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
32003803
Full Text :
https://doi.org/10.1093/europace/euaa006