Post-treatment analysis of irreversible electroporation waveforms delivered to human pancreatic cancer patients.

Authors :
Beitel-White N
Martin RCG
Davalos RV
Source :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2019 Jul; Vol. 2019, pp. 5518-5521.
Publication Year :
2019

Abstract

Irreversible electroporation (IRE) is a focal ablation therapy that uses high voltage, short electrical pulses to destroy tumor tissue. The success of treatment directly depends on exposure of the entire tumor to a lethal electric field magnitude. However, this exposure is difficult to predict ahead of time and it is challenging for clinicians to determine optimal treatment parameters. One method clinicians rely upon for the cessation of pulse delivery is to monitor the resistance value of the tissue, as the cells within the tissue will undergo changes during electroporation. This work presents a computational model which incorporates human pancreatic tumor conductivity, and compares predicted and measured output currents from IRE treatments of human patients. The measured currents vary widely from patient to patient, suggesting there may areas of high local conductivity in the treatment area.

Details

Language :
English
ISSN :
2694-0604
Volume :
2019
Database :
MEDLINE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Publication Type :
Academic Journal
Accession number :
31947104
Full Text :
https://doi.org/10.1109/EMBC.2019.8857259