Back to Search Start Over

Transvenous management of cardiac implantable electronic device late lead perforation.

Authors :
Jessel PM
Yadava M
Nazer B
Dewland TA
Miller J
Stecker EC
Bhamidipati CM
Song HK
Henrikson CA
Source :
Journal of cardiovascular electrophysiology [J Cardiovasc Electrophysiol] 2020 Feb; Vol. 31 (2), pp. 521-528. Date of Electronic Publication: 2020 Jan 10.
Publication Year :
2020

Abstract

Introduction: Late lead perforation (LLP), defined as perforation ā‰„30 days from cardiac implantable electronic device implant, is a rare diagnosis and little data exist regarding management practices and outcomes. The purpose of this study was to evaluate the occurrence, safety, and efficacy of transvenous management of clinically significant LLP.<br />Methods: The electronic medical records of a single-center tertiary hospital were reviewed for all patients who were referred for LLP or its sequelae.<br />Results: Eleven consecutive patients were identified from October 2011 to December 2018 with clinically significant LLP. Patients most often presented with pericardial symptoms with the exception of one asymptomatic patient. The median time from lead implant to intervention for LLP was 246 days. Nine patients were managed with an initial transvenous approach, with one requiring sternotomy (lead 6.3 years old). Two patients had a surgical approach, one performed at an outside hospital with subsequent death and another had a mini-thoracotomy, but the lead was removed percutaneously with no surgical repair. In this small cohort, there was no association between the lead extending beyond the parietal pericardium and surgical repair (Pā€‰=ā€‰.99).<br />Conclusion: Our single-center experience suggests that LLP can be initially managed with a cautious transvenous approach in most patients, but intraprocedural ultrasound for pericardial monitoring and a rescue plan with immediate surgical back up is mandatory.<br /> (© 2020 Wiley Periodicals, Inc.)

Details

Language :
English
ISSN :
1540-8167
Volume :
31
Issue :
2
Database :
MEDLINE
Journal :
Journal of cardiovascular electrophysiology
Publication Type :
Academic Journal
Accession number :
31908061
Full Text :
https://doi.org/10.1111/jce.14331