Validation of a custom spine biomechanics simulator: A case for standardization.
- Source :
-
Journal of biomechanics [J Biomech] 2020 Jan 02; Vol. 98, pp. 109470. Date of Electronic Publication: 2019 Nov 06. - Publication Year :
- 2020
-
Abstract
- Mechanical testing machines used in cadaveric spine biomechanics research vary between labs. It is a necessary first step to understand the capabilities and limitations in any testing machine prior to publishing experimental data. In this study, a reproducible protocol that uses a synthetic spine was developed and used to quantify the inherent rotation error and the ability to apply loads in a single physiologic plane (pure-moment) of a custom spine biomechanics simulator. Rotation error was evaluated by comparing data collected by the test machine and the data collected by an optical motion capture system. Pure-moment loading was assessed by comparing the out-of-plane loads to the primary plane load. Using synthetic functional spine units previously shown to have mechanics similar to the cadaveric human spine, the simulator was evaluated using a dynamic test protocol reflective of its future use in the study of cadaveric spine specimens. Rotation errors inherent in the test machine were <0.25° compared to motion capture. Out of plane loads were <4.0% of the primary plane load, which confirmed pure-moment loading. The authors suggest that a standard validation protocol for biomechanical spine testing machines is needed for transparency and accurate field-wide data interpretation and comparison. We offer recommendations based on the reproducible use of a synthetic spinal specimen for consideration.<br /> (Copyright © 2019 Elsevier Ltd. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-2380
- Volume :
- 98
- Database :
- MEDLINE
- Journal :
- Journal of biomechanics
- Publication Type :
- Academic Journal
- Accession number :
- 31740014
- Full Text :
- https://doi.org/10.1016/j.jbiomech.2019.109470