Back to Search
Start Over
Lock-in Amplifier-Based Impedance Detection of Tissue Type Using a Monopolar Injection Needle.
- Source :
-
Sensors (Basel, Switzerland) [Sensors (Basel)] 2019 Oct 23; Vol. 19 (21). Date of Electronic Publication: 2019 Oct 23. - Publication Year :
- 2019
-
Abstract
- For successful intra-articular injection therapy, it is essential to accurately position the tip of the injection needle into the target joint area while administering the drug into the affected tissue. In this study, we investigated the feasibility of a monopolar injection needle and lock-in amplifier (LIA)-based impedance measurement system for detecting the tissue type where the needle tip is located. After positioning the monopolar injection needle tip into the dermis, hypodermis, or muscle layer of pork tissue, the electrical impedance was measured in the frequency range of 10 Hz to 10 kHz. We observed a difference in the results based on the tissue type where the needle was positioned ( p -value < 0.01). Therefore, the monopolar injection needle with electrical impedance measurement can be used to improve intra-articular injection therapy through non-destructive and real-time monitoring of the needle position in the tissues.<br />Competing Interests: The author declares that there is no conflict of interest regarding the publication of this paper.
Details
- Language :
- English
- ISSN :
- 1424-8220
- Volume :
- 19
- Issue :
- 21
- Database :
- MEDLINE
- Journal :
- Sensors (Basel, Switzerland)
- Publication Type :
- Academic Journal
- Accession number :
- 31652819
- Full Text :
- https://doi.org/10.3390/s19214614