Back to Search
Start Over
Beta spectrum measurements using a quasi-4π detection system based on Si detectors.
- Source :
-
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2019 Dec; Vol. 154, pp. 108897. Date of Electronic Publication: 2019 Sep 17. - Publication Year :
- 2019
-
Abstract
- The accurate measurement of beta spectra is highly important in numerous fields such as nuclear energy, nuclear medicine, ionizing radiation metrology and fundamental physics. We have developed a beta spectrometer close to 4π configuration based on silicon detectors. The influence of self-absorption has been studied by means of Monte Carlo simulations and the source preparation technique has been optimized consequently. The measured spectra from <superscript>133</superscript> Ba and <superscript>36</superscript> Cl decays have been compared with PENELOPE simulations.<br /> (Copyright © 2019 Elsevier Ltd. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1872-9800
- Volume :
- 154
- Database :
- MEDLINE
- Journal :
- Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- Publication Type :
- Academic Journal
- Accession number :
- 31581058
- Full Text :
- https://doi.org/10.1016/j.apradiso.2019.108897