Back to Search
Start Over
The Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure.
- Source :
-
Spine [Spine (Phila Pa 1976)] 2020 Mar 15; Vol. 45 (6), pp. E346-E351. - Publication Year :
- 2020
-
Abstract
- Study Design: Case report.<br />Objective: To report a unique mechanical failure of magnetically controlled growing rods (MCGRs) that is related to continuous rod slippage.<br />Summary of Background Data: Despite the growing popularity of MCGR in the management of patients with early onset scoliosis, postoperative complications and reoperations are not uncommon. Unique complications or unplanned reoperations are observed in patients with MCGRs that are not seen in traditional growing rods. The complications include rod slippage, mismatch between targeted and achieved distraction length, metallosis, and actuator pin fracture. We have identified an unreported failure mechanism whereby deformed threads occur in the internal distraction mechanism of the MCGR. This phenomenon may indicate increased MCGR internal screw friction and increased distraction resistance. Increased friction and resistance ultimately led to distraction failure and revision surgery.<br />Methods: A girl with early onset scoliosis (EOS) was treated by dual MCGRs. She developed proximal junctional kyphosis (PJK) and continuous rod slippage. Distractions in the out-patient clinic and under sedation did not achieve successful rod lengthening. A tell-tale "crooked rod" radiological sign was identified by angulation between the actuator and the extendable portion of the rod.<br />Results: Rod exchange was performed and she is now 2 years after revision surgery with successful lengthening episodes. Upon review of the extracted MCGR, distraction was not possible even after rod removal and was only resumed after the screw was manually reinserted along its correct threads.<br />Conclusion: Deformed threads is a complication that cannot be reverted without rod removal and exchange. Users should be aware of this potential failure mechanism and proceed to early rod exchange rather than attempting any further distractions.<br />Level of Evidence: 4.
- Subjects :
- Bone Lengthening methods
Child, Preschool
Female
Humans
Kyphosis diagnostic imaging
Postoperative Complications diagnostic imaging
Postoperative Complications etiology
Prosthesis Failure adverse effects
Prosthesis Failure etiology
Reoperation methods
Scoliosis diagnostic imaging
Bone Lengthening adverse effects
Bone Screws adverse effects
Kyphosis surgery
Magnetic Phenomena
Postoperative Complications surgery
Scoliosis surgery
Subjects
Details
- Language :
- English
- ISSN :
- 1528-1159
- Volume :
- 45
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Spine
- Publication Type :
- Academic Journal
- Accession number :
- 31574056
- Full Text :
- https://doi.org/10.1097/BRS.0000000000003268