- Source :
-
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists [J Endovasc Ther] 2019 Dec; Vol. 26 (6), pp. 865-870. Date of Electronic Publication: 2019 Aug 28. - Publication Year :
- 2019
-
Abstract
- Purpose: To compare balloon-expandable covered stents (BECS) available for use as renal chimney grafts to the Advanta V12 approved for use with the Endurant endograft. Materials and Methods: A silicone model was manufactured based on preoperative computed tomography angiography (CTA) scans of a patient with a 7-cm juxtarenal aneurysm treated with an Endurant bifurcated endograft and an Advanta V12 covered stent for the 6-mm-diameter right renal artery. The model was placed in a flow box filled with a 37°C blood-mimicking solution equipped with a pulsatile pumping system (140/60 mm Hg). The tested BECS were the Advanta V12, the VBX, and the BeGraft+. A 36-mm-diameter Endurant II endograft with a 16-mm-diameter contralateral limb and a 6×59-mm BECS chimney graft were used in 9 consecutive tests (3 for each combination). After each implantation, the model was placed in the CT scanner, and 2 radiologists blinded to the test device independently measured the gutter areas and the patent chimney graft lumen at the level of the Endurant's suprarenal stent, at the level of the chimney's maximum curvature, and 10 mm inside the renal artery. The intraclass correlation coefficients (ICC) were calculated to assess interreader reliability. Results: The mean gutter areas were 19.3±7, 20.2±8, and 22.3±8 mm <superscript>2</superscript> for the Advanta, VBX, and BeGraft+, respectively (all p>0.05). At the level of the aortic endograft's suprarenal stent struts, the mean diameter of the Advanta V12, VBX, and BeGraft+ were 4.46±0.3, 4.12±0.4, and 4.12±0.3 mm, respectively (all p>0.05). At the level of the maximum chimney graft angulation, the mean diameters were 3.77±0.3, 4.16±0.1, and 3.74±0.3 mm, respectively (all p>0.05). In the right renal artery, the mean diameters were 3.91±0.2, 4.05±0.9, and 4.3±0.4 mm, respectively (all p>0.05). The ICCs varied between 0.7 and 0.9, indicating good agreement between readers. Conclusion: These in vitro findings showed comparable results between the Advanta V12 and the available BECS used in conjunction with the Endurant endograft according to the instructions for use. Further clinical evaluation is needed to confirm these results.
- Subjects :
- Angioplasty, Balloon adverse effects
Aorta, Abdominal diagnostic imaging
Aorta, Abdominal physiopathology
Aortic Aneurysm, Abdominal diagnostic imaging
Aortic Aneurysm, Abdominal physiopathology
Blood Vessel Prosthesis Implantation adverse effects
Hemodynamics
Humans
Models, Anatomic
Models, Cardiovascular
Prosthesis Design
Renal Artery diagnostic imaging
Renal Artery physiopathology
Renal Circulation
Angioplasty, Balloon instrumentation
Aorta, Abdominal surgery
Aortic Aneurysm, Abdominal surgery
Blood Vessel Prosthesis
Blood Vessel Prosthesis Implantation instrumentation
Renal Artery surgery
Stents
Subjects
Details
- Language :
- English
- ISSN :
- 1545-1550
- Volume :
- 26
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists
- Publication Type :
- Academic Journal
- Accession number :
- 31455137
- Full Text :
- https://doi.org/10.1177/1526602819872498