High resolution radiochromic film dosimetry: Comparison of a microdensitometer and an optical microscope.

Authors :
Pellicioli P
Bartzsch S
Donzelli M
Krisch M
Bräuer-Krisch E
Source :
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2019 Sep; Vol. 65, pp. 106-113. Date of Electronic Publication: 2019 Aug 23.
Publication Year :
2019

Abstract

Purpose: Microbeam radiation therapy is a developing technique that promises superior tumour control and better normal tissue tolerance using spatially fractionated X-ray beams only tens of micrometres wide. Radiochromic film dosimetry at micrometric scale was performed using a microdensitometer, but this instrument presents limitations in accuracy and precision, therefore the use of a microscope is suggested as alternative. The detailed procedures developed to use the two devices are reported allowing a comparison.<br />Methods: Films were irradiated with single microbeams and with arrays of 50 µm wide microbeams spaced by a 400 µm pitch, using a polychromatic beam with mean energy of 100 keV. The film dose measurements were performed using two independent instruments: a microdensitometer (MDM) and an optical microscope (OM).<br />Results: The mean values of the absolute dose measured with the two instruments differ by less than 5% but the OM provides reproducibility with a standard deviation of 1.2% compared to up to 7% for the MDM. The resolution of the OM was determined to be ~ 1 to 2 µm in both planar directions able to resolve pencil beams irradiation, while the MDM reaches at the best 20 µm resolution along scanning direction. The uncertainties related to the data acquisition are 2.5-3% for the OM and 9-15% for the MDM.<br />Conclusion: The comparison between the two devices validates that the OM provides equivalent results to the MDM with better precision, reproducibility and resolution. In addition, the possibility to study dose distributions in two-dimensions over wider areas definitely sanctions the OM as substitute of the MDM.<br /> (Copyright © 2019 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1724-191X
Volume :
65
Database :
MEDLINE
Journal :
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
Publication Type :
Academic Journal
Accession number :
31450120
Full Text :
https://doi.org/10.1016/j.ejmp.2019.08.012