Back to Search
Start Over
Use of 65 cm large caliber Dryseal sheaths to facilitate delivery of the Edwards SAPIEN valve to dysfunctional right ventricular outflow tracts.
- Source :
-
Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions [Catheter Cardiovasc Interv] 2019 Sep 01; Vol. 94 (3), pp. 409-413. Date of Electronic Publication: 2019 Aug 13. - Publication Year :
- 2019
-
Abstract
- Background: The Edwards SAPIEN valve and its delivery system may complicate transit through the right heart during transcatheter pulmonary valve replacement (tPVR). We report our early experience using a large diameter, 65 cm delivery sheath to facilitate delivery of the SAPIEN valve to the right ventricular outflow tract (RVOT).<br />Methods: Retrospective analysis of all patients from three large congenital heart centers undergoing tPVR with the Edwards SAPIEN valve delivered with the 65 cm Gore Dryseal Sheath.<br />Results: Over a 12 month period, 30 patients (17 female) with median age 17.5 years (range 8-72) underwent attempted tPVR with the SAPIEN valve delivered using the 65 cm Dryseal sheath (20-26Fr). All procedures resulted in successful valve delivery to the target area. Twenty patients had a native RVOT. The most commonly used valve diameter was 29 mm (n = 15) with the majority of cases requiring a 26Fr Dryseal sheath (n = 20). One patient with severe RVOT stenosis underwent prestenting. Median procedure time was 100 min (59-225). No patient had increase in tricuspid valve regurgitation as a consequence of valve delivery. One patient required a synchronous cardioversion for intraprocedural VT and another required ECMO postprocedure due to severe pre-existing left ventricular dysfunction. On median follow-up of 5 months, all patients had mild or less pulmonary regurgitation. Median peak Doppler velocity across the pulmonary valve was 2.2 m/s (1.7-4). There were no clinically relevant complications relating to vascular access.<br />Conclusions: Using 65 cm Dryseal sheaths facilitates delivery of SAPIEN valves in patients with dysfunctional RVOTs.<br /> (© 2019 Wiley Periodicals, Inc.)
- Subjects :
- Adolescent
Adult
Aged
Cardiac Catheterization adverse effects
Child
Colorado
Female
Heart Valve Prosthesis Implantation adverse effects
Hemodynamics
Humans
Ireland
Italy
Male
Middle Aged
Pulmonary Valve diagnostic imaging
Pulmonary Valve physiopathology
Pulmonary Valve Insufficiency diagnostic imaging
Pulmonary Valve Insufficiency physiopathology
Pulmonary Valve Stenosis diagnostic imaging
Pulmonary Valve Stenosis physiopathology
Recovery of Function
Retrospective Studies
Treatment Outcome
Young Adult
Cardiac Catheterization instrumentation
Cardiac Catheters
Heart Valve Prosthesis Implantation instrumentation
Pulmonary Valve surgery
Pulmonary Valve Insufficiency surgery
Pulmonary Valve Stenosis surgery
Stents
Subjects
Details
- Language :
- English
- ISSN :
- 1522-726X
- Volume :
- 94
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
- Publication Type :
- Academic Journal
- Accession number :
- 31408262
- Full Text :
- https://doi.org/10.1002/ccd.28409