Fibrotic Encapsulation Is the Dominant Source of Continuous Glucose Monitor Delays.

Authors :
McClatchey PM
McClain ES
Williams IM
Malabanan CM
James FD
Lord PC
Gregory JM
Cliffel DE
Wasserman DH
Source :
Diabetes [Diabetes] 2019 Oct; Vol. 68 (10), pp. 1892-1901. Date of Electronic Publication: 2019 Aug 09.
Publication Year :
2019

Abstract

Continuous glucose monitor (CGM) readings are delayed relative to blood glucose, and this delay is usually attributed to the latency of interstitial glucose levels. However, CGM-independent data suggest rapid equilibration of interstitial glucose. This study sought to determine the loci of CGM delays. Electrical current was measured directly from CGM electrodes to define sensor kinetics in the absence of smoothing algorithms. CGMs were implanted in mice, and sensor versus blood glucose responses were measured after an intravenous glucose challenge. Dispersion of a fluorescent glucose analog (2-NBDG) into the CGM microenvironment was observed in vivo using intravital microscopy. Tissue deposited on the sensor and nonimplanted subcutaneous adipose tissue was then collected for histological analysis. The time to half-maximum CGM response in vitro was 35 ± 2 s. In vivo, CGMs took 24 ± 7 min to reach maximum current versus 2 ± 1 min to maximum blood glucose ( P = 0.0017). 2-NBDG took 21 ± 7 min to reach maximum fluorescence at the sensor versus 6 ± 6 min in adipose tissue ( P = 0.0011). Collagen content was closely correlated with 2-NBDG latency ( R = 0.96, P = 0.0004). Diffusion of glucose into the tissue deposited on a CGM is substantially delayed relative to interstitial fluid. A CGM that resists fibrous encapsulation would better approximate real-time deviations in blood glucose.<br /> (© 2019 by the American Diabetes Association.)

Details

Language :
English
ISSN :
1939-327X
Volume :
68
Issue :
10
Database :
MEDLINE
Journal :
Diabetes
Publication Type :
Academic Journal
Accession number :
31399432
Full Text :
https://doi.org/10.2337/db19-0229