Back to Search Start Over

Improved Atomic Force Microscopy Stiffness Measurements of Nanoscale Liposomes by Cantilever Tip Shape Evaluation.

Authors :
Takechi-Haraya Y
Goda Y
Izutsu K
Sakai-Kato K
Source :
Analytical chemistry [Anal Chem] 2019 Aug 20; Vol. 91 (16), pp. 10432-10440. Date of Electronic Publication: 2019 Aug 07.
Publication Year :
2019

Abstract

The stiffness of nanoscale liposomes, as measured by atomic force microscopy (AFM), was investigated as a function of temperature, immobilization on solid substrates, and cantilever tip shape. The liposomes were composed of saturated lipids and cholesterol, and the stiffness values did not change over the temperature range of 25-37 °C and were independent of immobilization methods. However, the stiffness varied with the tip shape of the cantilever. Therefore, 24 cantilevers were evaluated in terms of tip shape and aspect ratio (length/width) via a nonblind tip reconstruction (NBTR) method that used a tip characterizer with isolated line structures having specified dimensions. A standard for screening the tip geometry was established. A 24-fold improvement in stiffness precision in terms of relative standard deviation was demonstrated by using at least three cantilevers that meet the criteria of having a tip aspect ratio greater than 2.5 and a quadratic tip shape function. A significant difference in stiffness was subsequently revealed between dipalmitoylphosphatidylcholine-cholesterol (1:1 molar ratio) and egg yolk phosphatidylcholine-cholesterol (1:1 molar ratio) liposomes. Tip analysis using NBTR improved the precision of AFM stiffness measurements, which will enable the control of mechanical properties of nanoscale liposomes for various applications.

Details

Language :
English
ISSN :
1520-6882
Volume :
91
Issue :
16
Database :
MEDLINE
Journal :
Analytical chemistry
Publication Type :
Academic Journal
Accession number :
31390864
Full Text :
https://doi.org/10.1021/acs.analchem.9b00250