Back to Search
Start Over
Glucose Control in Adults with Type 1 Diabetes Using a Medtronic Prototype Enhanced-Hybrid Closed-Loop System: A Feasibility Study.
- Source :
-
Diabetes technology & therapeutics [Diabetes Technol Ther] 2019 Sep; Vol. 21 (9), pp. 499-506. Date of Electronic Publication: 2019 Jul 02. - Publication Year :
- 2019
-
Abstract
- Background: Experience from first-generation closed-loop (CL) systems informs refinements to enhance glucose control and user acceptance. A next-generation prototype enhanced-hybrid CL (E-HCL) system incorporates iterative changes to the Medtronic MiniMed 670G CL system, including automated correction boluses, lower target glucose level, and user enhancements. The aim was to explore safety, system performance, and glucose control using E-HCL in adults with type 1 diabetes. Methods: Twelve adults underwent this first in-human feasibility study. After a 1-week run-in using open-loop (OL), E-HCL was activated at the start of a supervised 1-week hotel phase, followed by 3 weeks free living at home. Supervised challenges included two meal interventions (unannounced and late meal bolus) and a sensor calibration intervention. Primary outcome was sensor glucose time-in-range (TIR); OL run-in and E-HCL at home were compared by Wilcoxon signed-rank test. Results: Twelve adults (seven men; median [interquartile range] age 48 [39, 57] years; HbA <subscript>1c</subscript> 6.8 [6.2, 7.2]%, 51 [44, 55] mmol/mol; diabetes duration 31 [13, 41] years) completed the protocol. E-HCL resulted in greater TIR (85.3 [79.4, 88.4]% vs. 75.0 [66.6, 83.7]%, P = 0.003) and lower mean sensor glucose (123.0 [119.3, 129.6] mg/dL vs. 143.5 [135.8, 154.5] mg/dL, P = 0.002) than OL. Time spent <70 mg/dL increased using E-HCL (4.4 [3.3, 6.1]% vs. 3.0 [1.8, 3.8]%, P = 0.02) with no difference in time <54 mg/dL ( P = 0.64). Time in CL was 99.98 [99.0, 100.0]%. All participants were satisfied using E-HCL. Conclusions: In adults with well-controlled HbA <subscript>1c</subscript> levels, a prototype E-HCL resulted in high TIR, few CL exits, and positive user experiences at the expense of increased hypoglycemia (<70 mg/dL). E-HCL represents a positive step in the journey toward optimizing glucose control in people living with type 1 diabetes.
- Subjects :
- Adult
Calibration
Diabetes Mellitus, Type 1 complications
Diabetes Mellitus, Type 1 drug therapy
Feasibility Studies
Female
Glycated Hemoglobin analysis
Humans
Hypoglycemia epidemiology
Hypoglycemia etiology
Hypoglycemia prevention & control
Hypoglycemic Agents administration & dosage
Insulin administration & dosage
Male
Middle Aged
Time Factors
Treatment Outcome
Blood Glucose analysis
Blood Glucose Self-Monitoring instrumentation
Diabetes Mellitus, Type 1 blood
Insulin Infusion Systems
Subjects
Details
- Language :
- English
- ISSN :
- 1557-8593
- Volume :
- 21
- Issue :
- 9
- Database :
- MEDLINE
- Journal :
- Diabetes technology & therapeutics
- Publication Type :
- Academic Journal
- Accession number :
- 31264889
- Full Text :
- https://doi.org/10.1089/dia.2019.0120