Back to Search Start Over

Residual Negative Pressure in Vacuum Blood-Collection Tube and Hemolysis in Pediatric Blood Specimens.

Authors :
Hu J
Zhang QX
Xiao TT
Pan MC
Cai YM
Source :
Laboratory medicine [Lab Med] 2020 Jan 02; Vol. 51 (1), pp. 41-46.
Publication Year :
2020

Abstract

Objective: To determine a method to reduce specimen hemolysis rates in pediatric blood specimens.<br />Methods: A total of 290 blood specimens from pediatric patients were classified into the capped group or uncapped group. The hemolysis index and levels of lactate dehydrogenase (LDH) were measured using an automated biochemical analyzer. Also, we performed a paired test to measure the concentration of free hemoglobin in specimens from 25 randomly selected healthy adult volunteers, using a direct spectrophotometric technique.<br />Results: The hemolytic rate of capped specimens was 2-fold higher than that of uncapped specimens. We found significant differences for LDH. Also, there was a significant difference in the concentration of free hemoglobin in the random-volunteers test.<br />Conclusions: Eliminating the residual negative pressure of vacuum blood-collection tubes was effective at reducing the macrohemolysis and/or microhemolysis rate.<br /> (© American Society for Clinical Pathology 2019. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1943-7730
Volume :
51
Issue :
1
Database :
MEDLINE
Journal :
Laboratory medicine
Publication Type :
Academic Journal
Accession number :
31185079
Full Text :
https://doi.org/10.1093/labmed/lmz026