Biomechanical Analysis of Medial-Row All-Suture Suture Anchor Fixation for Rotator Cuff Repair in a Pair-Matched Cadaveric Model.

Authors :
Bernardoni ED
Frank RM
Veera SS
Waterman BR
Griffin JW
Shewman EF
Cole BJ
Romeo AA
Verma NN
Source :
Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association [Arthroscopy] 2019 May; Vol. 35 (5), pp. 1370-1376. Date of Electronic Publication: 2019 Apr 15.
Publication Year :
2019

Abstract

Purpose: To compare the biomechanical properties of all-suture suture anchors (ASSAs) with conventional suture anchors (CSAs) for double-row rotator cuff repair (RCR).<br />Methods: Fourteen fresh-frozen human cadaveric shoulders were randomized into 2 RCR treatment groups: ASSA and CSA. All constructs received a double-row repair, with the lateral-row implants consisting of two 5.5-mm PEEK (polyether ether ketone) Footprint anchors. Each construct was loaded to a 10-N preload for 2 minutes, followed by cyclic loading from 10 to 160 N at a rate of 100 N/s for 100 cycles. Load-to-failure testing was performed immediately after cyclic loading testing at 1 mm/s from the zero position until failure. Cyclic creep, elongation amplitude, maximum load, stiffness, energy, and failure mode were recorded.<br />Results: No significant difference in cyclic creep (P = .117) or elongation amplitude (P = .428) was found between the ASSA and CSA groups during cyclic testing. Three specimens in each group (43% in each) failed by the suture tearing through the tendon. The remaining specimens in each group failed by the anchor pulling out of the humeral head. The mean maximum load was 617.73 ± 177.77 N and 545.13 ± 212.98 N for the ASSA and CSA groups, respectively (P = .339). Maximum elongation before failure was not different between groups (P = .122). Mean energy and stiffness were not statistically different between the ASSA and CSA groups (P = .629 and P = .973, respectively).<br />Conclusions: In this cadaveric analysis with a simplified unidirectional experimental setup, failure mechanics and maximum load between the ASSA and CSA constructs were similar, with no difference in energy and stiffness. Although the ASSA group showed slightly larger elongation than the CSA group, these differences may not be clinically relevant.<br />Clinical Relevance: This study provides a biomechanical head-to-head comparison of ASSAs and CSAs, indicating that ASSAs may be clinically equivalent to CSAs for use in an RCR.<br /> (Copyright © 2019 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1526-3231
Volume :
35
Issue :
5
Database :
MEDLINE
Journal :
Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association
Publication Type :
Academic Journal
Accession number :
31000387
Full Text :
https://doi.org/10.1016/j.arthro.2019.01.023