Back to Search Start Over

Rapid direct determination of tin in beverages using energy dispersive X-ray fluorescence.

Authors :
Lerner N
Sedgi I
Chernia Z
Zeiri O
Source :
Talanta [Talanta] 2019 Jul 01; Vol. 199, pp. 662-666. Date of Electronic Publication: 2019 Mar 04.
Publication Year :
2019

Abstract

This article presents a new method for the simpler and faster determination of tin in beverages using EDXRF. Absorption coefficients for aqueous calibration samples were calculated and shown to be nearly identical to those of the beverage samples, thus permitting the use of aqueous standard solutions for external calibration. Beverage samples could then be measured directly using the external calibration. Determination of tin using this method takes 4 min. The LOD and LOQ were 4 mg L <superscript>-1</superscript> and 15 mg L <superscript>-1</superscript> respectively, and the precision was 3.89%. Different canned beverages (cold coffee, various fruit juices) were measured and the results compared to the concentrations obtained using ICP-OES after digestion. The two methods showed good compatibility, thus establishing the newly developed method as a rapid, simple, and accurate method for the determination of tin in beverages.<br /> (Copyright © 2019 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-3573
Volume :
199
Database :
MEDLINE
Journal :
Talanta
Publication Type :
Academic Journal
Accession number :
30952312
Full Text :
https://doi.org/10.1016/j.talanta.2019.03.024