Back to Search
Start Over
LMA® Protector™ versus traditional LMA to perform endobronchial ultrasound-guided transbronchial needle aspiration: a retrospective analysis.
- Source :
-
Minerva anestesiologica [Minerva Anestesiol] 2019 Jul; Vol. 85 (7), pp. 756-762. Date of Electronic Publication: 2019 Mar 29. - Publication Year :
- 2019
-
Abstract
- Background: The aim of this study was to evaluate the use of laryngeal mask airway (LMA)® Protector™ by comparison with traditional LMA for performing endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA).<br />Methods: This was a retrospective observational single-center study including 143 patients who underwent EBUS-TBNA for mediastinal staging of lung cancer. Patients were retrospectively divided into two groups based on whether a traditional LMA (traditional LMA group) or LMA Protector was used. Anesthesiologist outcomes, diagnostic yield of EBUS-TBNA, and complications related to the procedure were computed for each group and statistically compared.<br />Results: LMA traditional group and LMA Protector group counted 70 and 73 patients, respectively. LMA traditional group versus LMA Protector group showed no significant difference on time of LMA insertion (120±25 vs. 118±39 s; P=0.49), reposition rates (18% vs. 16%; P=0.78); systolic pressure (140±55 vs. 118±37 mmHg; P=0.59); diastolic pressure (82±15 vs. 90±26 mmHg; P=0.39); heart rate (82±9.9 vs. 83±20 bpm; P=0.49); SpO2 values (93±21% vs. 92±14%; P=0.63); diagnostic accuracy (91.3% vs. 92%; P=0.95), and patients' complications as nausea (4% vs. 3%; P=0.61); vomiting (3% vs. 1%, P=0.96); gastric aspiration (7% vs. 1%; P=0.08); and sore throat (7% vs. 3%; P=0.22). Conversely, LMA traditional group versus LMA Protector group presented a longer procedural time (47±23 vs. 38±17 s; P=0.02), higher number of passage to biopsy target lesion (4±0.5 vs. 3.1±0.6; P=0.01); higher rate of balloon ultrasound rupture (11% vs. 1%; P=0.01).<br />Conclusions: EBUS-TBNA conducted with LMA Protector is a useful strategy that reduced the procedural time and in theory ensured the comfort of patients. Our results should be confirmed by larger, prospective, randomized studies.
- Subjects :
- Aged
Biopsy, Needle adverse effects
Computer Systems
Endosonography adverse effects
Equipment Design
Female
Hemodynamics
Humans
Lung Neoplasms pathology
Male
Nausea etiology
Neoplasm Staging methods
Pharynx injuries
Respiratory Aspiration of Gastric Contents etiology
Retrospective Studies
Vomiting etiology
Biopsy, Needle methods
Bronchi
Endosonography methods
Image-Guided Biopsy methods
Laryngeal Masks adverse effects
Subjects
Details
- Language :
- English
- ISSN :
- 1827-1596
- Volume :
- 85
- Issue :
- 7
- Database :
- MEDLINE
- Journal :
- Minerva anestesiologica
- Publication Type :
- Academic Journal
- Accession number :
- 30938120
- Full Text :
- https://doi.org/10.23736/S0375-9393.19.13328-7