Back to Search Start Over

MEASUREMENT OF OPERATIONAL QUANTITIES Hp(0.07) AND Hp(3) FOR INDIGENOUSLY DEVELOPED 106Ru/106Rh SOURCE USING AN EXTRAPOLATION CHAMBER.

Authors :
Vandana S
Bakshi AK
Behrens R
Chattaraj A
Saxena SK
Dhami PS
Panja S
Jagasia P
Selvam TP
Dash A
Datta D
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2019 Dec 31; Vol. 185 (3), pp. 376-386.
Publication Year :
2019

Abstract

In the present study, a prototype 106Ru/106Rh source was fabricated using high level liquid waste from reactor fuel, fixed in a stainless steel housing with a window and backing made of silver. The study involves measurement of the operational quantities Hp(0.07), Hp(3) and the percentage depth dose (PDD) using an extrapolation chamber. It also involves determination of necessary correction factors to arrive at Hp(0.07) and Hp(3) following International Organisation for Standardisation (ISO) and methods suggested in literature. The study facilitates incorporation of the 106Ru/106Rh source as a beta reference source for quality assurance programme in TLD personnel monitoring as per the guidelines of ISO.<br /> (© The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1742-3406
Volume :
185
Issue :
3
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
30916771
Full Text :
https://doi.org/10.1093/rpd/ncz024