Back to Search
Start Over
Hands-on defibrillation with a safety barrier: An analysis of potential risk to rescuers.
- Source :
-
Resuscitation [Resuscitation] 2019 May; Vol. 138, pp. 110-113. Date of Electronic Publication: 2019 Mar 09. - Publication Year :
- 2019
-
Abstract
- Background: Interruptions in compressions reducethe efficacy of cardiopulmonary resuscitation (CPR) and are inevitable during hands-off periods for shocks. Clinical exam gloves were found to facilitate safe contact with patients during shock delivery but the safety of this practice has been questioned. Polyethylene is of interest because of its safety record in the medical arena and its electrical insulation properties.<br />Methods: This study measured the current leak through 2 mil (0.002 inch) polyethylene drapes during shock delivery. The current leak was assessed by measurement of voltage changes in a circuit recommended by the International Electrotechnical Commission (IEC) for current leak safety testing. Current flowed off the drape, through the circuit and to electric ground in a manner consistent with standardized testing. Perceptibility was assessed in a subset with the investigator's bare hands pressed into the drape during shock delivery.<br />Results: Thirty-three patients undergoing elective cardioversion at Emory University Hospital underwent analysis (age 23-90, 36% female). Biphasic energies were 200-360 J. The root mean square (RMS) current leak averaged 0.072 ± 0.022 mA and peak current leak averaged 0.67 ± 0.21 which is well below IEC recommendations of 3.5 mA RMS and 5.0 mA peak. Finally, no instances of dielectric breakdown occurred and no shocks were perceptible.<br />Conclusions: Polyethylene is a common medical material which may facilitate safe hands-on defibrillation. Our data illustrates that a thin, semitransparent layer of polyethylene is a safe and feasible adjunct to cardiac arrest kits to allow continued compressions and simplification of the CPR process.<br /> (Copyright © 2019 Elsevier B.V. All rights reserved.)
- Subjects :
- Accidents, Occupational prevention & control
Combined Modality Therapy methods
Defibrillators adverse effects
Electric Conductivity
Humans
Materials Testing methods
Polyethylene
Cardiopulmonary Resuscitation methods
Electric Countershock adverse effects
Electric Countershock methods
Electric Injuries etiology
Electric Injuries prevention & control
Gloves, Surgical standards
Heart Arrest therapy
Heart Massage methods
Safety
Subjects
Details
- Language :
- English
- ISSN :
- 1873-1570
- Volume :
- 138
- Database :
- MEDLINE
- Journal :
- Resuscitation
- Publication Type :
- Academic Journal
- Accession number :
- 30862529
- Full Text :
- https://doi.org/10.1016/j.resuscitation.2019.02.043