SPARKLING: variable-density k-space filling curves for accelerated T 2 * -weighted MRI.
- Source :
-
Magnetic resonance in medicine [Magn Reson Med] 2019 Jun; Vol. 81 (6), pp. 3643-3661. Date of Electronic Publication: 2019 Feb 17. - Publication Year :
- 2019
-
Abstract
- Purpose: To present a new optimition-driven design of optimal k-space trajectories in the context of compressed sensing: Spreading Projection Algorithm for Rapid K-space sampLING (SPARKLING).<br />Theory: The SPARKLING algorithm is a versatile method inspired from stippling techniques that automatically generates optimized sampling patterns compatible with MR hardware constraints on maximum gradient amplitude and slew rate. These non-Cartesian sampling curves are designed to comply with key criteria for optimal sampling: a controlled distribution of samples (e.g., variable density) and a locally uniform k-space coverage.<br />Methods: Ex vivo and in vivo prospective T 2 * -weighted acquisitions were performed on a 7-Tesla scanner using the SPARKLING trajectories for various setups and target densities. Our method was compared to radial and variable-density spiral trajectories for high-resolution imaging.<br />Results: Combining sampling efficiency with compressed sensing, the proposed sampling patterns allowed up to 20-fold reductions in MR scan time (compared to fully sampled Cartesian acquisitions) for two-dimensional T 2 * -weighted imaging without deterioration of image quality, as demonstrated by our experimental results at 7 Tesla on in vivo human brains for a high in-plane resolution of 390 μm. In comparison to existing non-Cartesian sampling strategies, the proposed technique also yielded superior image quality.<br />Conclusions: The proposed optimization-driven design of k-space trajectories is a versatile framework that is able to enhance MR sampling performance in the context of compressed sensing.<br /> (© 2019 International Society for Magnetic Resonance in Medicine.)
Details
- Language :
- English
- ISSN :
- 1522-2594
- Volume :
- 81
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Magnetic resonance in medicine
- Publication Type :
- Academic Journal
- Accession number :
- 30773679
- Full Text :
- https://doi.org/10.1002/mrm.27678