Back to Search
Start Over
Susceptibility of Cardiac Troponin Assays to Biotin Interference.
- Source :
-
American journal of clinical pathology [Am J Clin Pathol] 2019 Apr 02; Vol. 151 (5), pp. 486-493. - Publication Year :
- 2019
-
Abstract
- Objectives: To investigate biotin interference on three cardiac troponin (cTn) assays and demonstrate a method to overcome biotin interference.<br />Methods: cTn levels were measured in (1) plasma from healthy volunteers on 10-mg daily biotin supplementation mixed with a plasma with known elevated troponin, (2) plasmas with known elevated cTn after mixing in reagent biotin to simulate supplementation, and (3) biotin-spiked plasma specimens pretreated with streptavidin-agarose beads.<br />Results: Daily biotin ingestion (10 mg) and studies simulating daily biotin use resulted in significant interference in the Gen5 cardiac troponin T (cTnT) assay; the contemporary Gen 4 cTnT and high-sensitivity cardiac troponin I (hs-cTnI) assays were unaffected. The biotin interference threshold was 31, 315, and more than 2,000 ng/mL for Gen5 cTnT, cTnT, and hs-cTnI assays, respectively. Streptavidin pretreatment blocked biotin interference in cTn assays.<br />Conclusions: Biotin interference is possible at plasma concentrations achievable by ingestion of over-the-counter supplements that may lead to delayed or missed diagnosis of myocardial injury with the Gen5 cTnT assay.<br /> (© American Society for Clinical Pathology, 2019. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.)
Details
- Language :
- English
- ISSN :
- 1943-7722
- Volume :
- 151
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- American journal of clinical pathology
- Publication Type :
- Academic Journal
- Accession number :
- 30715102
- Full Text :
- https://doi.org/10.1093/ajcp/aqy172