Back to Search
Start Over
Pre-Use Ureteroscope Contamination after High Level Disinfection: Reprocessing Effectiveness and the Relation with Cumulative Ureteroscope Use.
- Source :
-
The Journal of urology [J Urol] 2019 Jun; Vol. 201 (6), pp. 1144-1151. - Publication Year :
- 2019
-
Abstract
- Purpose: We assessed the frequency of preoperative and persistent microbial contamination of flexible ureteroscopes after reprocessing and the relation of contamination to cumulative ureteroscope use.<br />Materials and Methods: We evaluated the effectiveness of high level disinfection with peracetic acid as well as data on ureteroscope use for 20 new flexible ureteroscopes from December 2015 to December 2017 at a single center. In the operating room pre-use and postuse microbial samples of the ureteroscope shaft and working channel were collected to evaluate microbial contamination after reprocessing. Positive cultures were defined as 30 cfu/ml or greater of skin flora, or 10 cfu/ml or greater of uropathogenic microorganisms. A generalized estimating equation model was used to analyze whether cumulative ureteroscope use was associated with positive pre-use cultures.<br />Results: Microbial samples were collected during 389 procedures. Pre-use ureteroscope cultures were positive in 47 of 389 procedures (12.1%), of which uropathogens were found in 9 of 389 (2.3%) and skin flora in 38 of 389 (9.8%). Urinary tract infection symptoms did not develop in any of the patients who underwent surgery with a uropathogen contaminated ureteroscope. In 1 case the pre-use culture contained the same bacteria type as the prior postuse culture. Cumulative ureteroscope use was not associated with a higher probability of positive cultures.<br />Conclusions: Microbial contamination of reprocessed ureteroscopes was found in an eighth of all procedures. Notably uropathogenic microorganisms were discovered in a small proportion of all procedures. Persistent ureteroscope contamination with uropathogens was only rarely encountered. Cumulative ureteroscope use was not associated with a higher probability of microbial contamination.
- Subjects :
- Preoperative Period
Prospective Studies
Ureteroscopes statistics & numerical data
Disinfectants
Disinfection methods
Equipment Contamination
Equipment Reuse statistics & numerical data
Equipment and Supplies Utilization statistics & numerical data
Peracetic Acid
Ureteroscopes microbiology
Subjects
Details
- Language :
- English
- ISSN :
- 1527-3792
- Volume :
- 201
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- The Journal of urology
- Publication Type :
- Academic Journal
- Accession number :
- 30707130
- Full Text :
- https://doi.org/10.1097/JU.0000000000000108