Back to Search
Start Over
Systematic overestimation of human serum albumin by capillary zone electrophoresis method due to monoclonal immunoglobulin interferences.
- Source :
-
Clinica chimica acta; international journal of clinical chemistry [Clin Chim Acta] 2019 Apr; Vol. 491, pp. 74-80. Date of Electronic Publication: 2019 Jan 11. - Publication Year :
- 2019
-
Abstract
- Background: The capillary zone electrophoresis method of albumin measurement is frequently used in monoclonal gammopathy patients but some studies suggest poor performances of the method in this population. The aim of this study was to analyse the impact of serum monoclonal immunoglobulins on human serum albumin determination by capillary zone electrophoresis method compared to other available methods.<br />Method: We prospectively measured albumin in 100 freshly collected non-frozen serum samples in a monoclonal gammopathy patients population, by using four different methods: the capillary zone electrophoresis method, the bromocresol purple dye method, the nephelometric method and the turbidimetric method. Differences in albumin values between the different methods were analysed with respect to serum monoclonal immunoglobulin concentration. These differences were further investigated by measuring albumin levels in human serum samples spiked with exogenous monoclonal immunoglobulins.<br />Results: Human serum albumin difference values between capillary zone electrophoresis compared to immunonephelometry method are significantly correlated with increasing monoclonal immunoglobulins concentrations: regression analyses revealed a correlation coefficient r <superscript>2</superscript> = 0.60 and a slope of 0.14 (0.12-0.17, 95% confidence interval). The capillary zone electrophoresis method overestimated serum albumin levels by up to 67% (12 g/L) when monoclonal immunoglobulin level was 63 g/L. The determination of albumin levels in human serum samples spiked with exogenous monoclonal immunoglobulins showed an overestimation of human serum albumin measurement by the capillary zone electrophoresis method proportional to the amount of monoclonal immunoglobulin added in the serum with a slope of 0.19 (0.18-0.20, 95% confidence interval).<br />Conclusion: Monoclonal immunoglobulins directly interfere with serum albumin measurement by the capillary zone electrophoresis method leading to a systematic overestimation of serum albumin concentrations proportional to the serum monoclonal immunoglobulin level.<br /> (Copyright © 2019 Elsevier B.V. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-3492
- Volume :
- 491
- Database :
- MEDLINE
- Journal :
- Clinica chimica acta; international journal of clinical chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 30641058
- Full Text :
- https://doi.org/10.1016/j.cca.2019.01.010