INVESTIGATION OF APPLICABILITY OF PURE PROPANE GAS FOR MICRODOSIMETRY AT NEUTRON FIELDS: A MONTE CARLO STUDY.

Authors :
Chattaraj A
Selvam TP
Datta D
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2019 Nov 30; Vol. 185 (1), pp. 74-86.
Publication Year :
2019

Abstract

Applicability of pure propane gas for microdosimetric measurements in neutron fields was investigated using the FLUKA Monte Carlo code. Monoenergetic neutrons in the energy range 1 keV-20 MeV and the ISO-neutron sources such as 241Am-Be, 241Am-B, 252Cf and 252Cf + D2O were considered in the present study. The tissue-equivalent proportional counter (TEPC) simulated in the study was LET-1/2 (by Far West Technology) with site sizes 1, 2 and 8 μm. The study demonstrates that for a given site size, the TEPC filled with tissue-equivalent propane and pure propane gases produce similar microdosimetric distributions when the density of pure propane gas is lowered appropriately. For the ISO-neutron sources, the density of propane gas requires scaling by a factor 0.85. For the monoenergetic neutrons, depending upon the neutron energy, the values of scaling factors are in the range of 0.58-0.93.<br /> (© The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)