Back to Search Start Over

Simultaneous LIBS signal and plasma density measurement for quantitative insight into signal instability at elevated pressure.

Authors :
Patnaik AK
Wu Y
Hsu PS
Gragston M
Zhang Z
Gord JR
Roy S
Source :
Optics express [Opt Express] 2018 Oct 01; Vol. 26 (20), pp. 25750-25760.
Publication Year :
2018

Abstract

Laser-induced breakdown spectroscopy (LIBS) evaluates the emission spectra of ions, radicals, and atoms generated from the breakdown of molecules by the incident laser; however, the LIBS signal is unstable at elevated pressures. To understand the cause of the signal instability, we perform simultaneous time-resolved measurements of the electron density and LIBS emission signal for nitrogen (568 nm) and hydrogen (656 nm) at high pressure (up to 11 bars). From correlations between the LIBS signal and electron number density, we find that the uncontrollable generation of excess electrons at high pressure causes high instability in the high-pressure LIBS signal. A possible method using ultrafast lasers is proposed to circumvent the uncontrolled electron generation and improve signal stability at high pressure.

Details

Language :
English
ISSN :
1094-4087
Volume :
26
Issue :
20
Database :
MEDLINE
Journal :
Optics express
Publication Type :
Academic Journal
Accession number :
30469671
Full Text :
https://doi.org/10.1364/OE.26.025750