Back to Search
Start Over
Clinical Performance of the BioPlex 2200 Syphilis Total & RPR Assay at a Tertiary Medical Center with a High Rate of Syphilis.
- Source :
-
Journal of clinical microbiology [J Clin Microbiol] 2019 Jan 02; Vol. 57 (1). Date of Electronic Publication: 2019 Jan 02 (Print Publication: 2019). - Publication Year :
- 2019
-
Abstract
- Manual treponemal and nontreponemal serologic testing has historically been used for the diagnosis of syphilis. This approach is simple and reproducible but labor intensive. Recently, the FDA cleared the fully automated BioPlex 2200 Syphilis Total & RPR assay for the detection of treponemal and nontreponemal antibodies. We evaluated the clinical performance of this assay at a tertiary medical center with a high syphilis prevalence. Prospective consecutively collected ( n = 400) and known RPR-positive ( n = 100) specimens were compared using predicate manual rapid plasma reagin (RPR) and fluorescent treponemal antibody absorption (FTA) methods and the BioPlex 2200 Syphilis Total & RPR assay. Positive and negative percent agreements (PPA and NPA, respectively) between the assays were calculated. The PPA and NPA between the manual and BioPlex 2200 RPR results for the prospective population were 85% (17/20; 95% confidence interval [CI], 69% to 100%) and 98% (373/380; 95% CI, 97% to 99%), respectively. The PPA for the manual RPR-positive population was 88% (88/100; 95% CI, 82% to 94%). Overall, the manual and BioPlex 2200 RPR titers demonstrated 78% (99/127) concordance within ±1 dilution and 94% (120/127) within ±2 dilutions. An interpretation of the syphilis serologic profile using the traditional algorithm showed a concordance of 99.5% in the prospective population and 85% in the manual RPR-positive cohort. The performance of the BioPlex 2200 Syphilis Total & RPR assay is comparable to those of manual methods. The high NPA of this assay combined with the ability to automate a historically labor-intensive assay is an appealing attribute for syphilis screening in a high-volume laboratory.<br /> (Copyright © 2019 American Society for Microbiology.)
- Subjects :
- Adolescent
Adult
Aged
Aged, 80 and over
Algorithms
Antibodies, Bacterial blood
Automation, Laboratory
Child
Child, Preschool
Female
Humans
Infant
Infant, Newborn
Male
Middle Aged
Prospective Studies
Reagent Kits, Diagnostic
Reagins blood
Syphilis blood
Syphilis microbiology
Tertiary Care Centers
Treponema immunology
Young Adult
Immunoenzyme Techniques
Mass Screening methods
Syphilis diagnosis
Syphilis Serodiagnosis methods
Treponema isolation & purification
Subjects
Details
- Language :
- English
- ISSN :
- 1098-660X
- Volume :
- 57
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of clinical microbiology
- Publication Type :
- Academic Journal
- Accession number :
- 30429251
- Full Text :
- https://doi.org/10.1128/JCM.01487-18