Back to Search Start Over

Development of a multi-channel capacitive probe for electric field measurements with fine spatial and high time resolution.

Authors :
Nishizawa T
Almagri AF
Goodman W
Ohshima S
Sarff JS
Source :
The Review of scientific instruments [Rev Sci Instrum] 2018 Oct; Vol. 89 (10), pp. 10J118.
Publication Year :
2018

Abstract

A capacitive probe [Tan et al. , Rev. Sci. Instrum. 88 , 023502 (2017)] is one of a few diagnostics that is directly sensitive to the plasma potential. Using this diagnostic technique, a Multi-channel Linear Capacitive Probe (MLCP) is developed for turbulence measurements. The MLCP has 10 spatial channels and provides 9 points of radial electric field measurements simultaneously with a spatial step of 7 mm. A new readout circuit and a correction technique for low frequency attenuation are also developed to achieve the required spatial and time resolution. A performance test of the MLCP using a reversed field pinch plasma confirms that the MLCP resolves sub-centimeter structures of the equilibrium radial electric field profile and fluctuations up to 680 kHz.

Details

Language :
English
ISSN :
1089-7623
Volume :
89
Issue :
10
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
30399956