Back to Search Start Over

Development of a high-fidelity minimally invasive mitral valve surgery simulator.

Authors :
Sardari Nia P
Daemen JHT
Maessen JG
Source :
The Journal of thoracic and cardiovascular surgery [J Thorac Cardiovasc Surg] 2019 Apr; Vol. 157 (4), pp. 1567-1574. Date of Electronic Publication: 2018 Sep 29.
Publication Year :
2019

Abstract

Objectives: The aim of this study was to develop a high-fidelity minimally invasive mitral valve surgery (MIMVS) simulator.<br />Methods: The process of industrial serial design was applied based on pre-set requirements, acquired by interviewing experienced mitral surgeons. A thoracic torso with endoscopic and robotic access and disposable silicone mitral valve apparatus with a feedback system was developed. The feedback system was based on 4 cameras around the silicone valve and an edge detection algorithm to calculate suture depth and width. Validity of simulator measurements was assessed by comparing simulator-generated values with measurements done manually on 3-dimensional reconstructed micro-computed tomography scan of the same sutures. Independent surgeons tested the simulator between 2014 and 2018, whereupon an evaluation was done through a questionnaire.<br />Results: The feedback system was able to provide width and depth measurements, which were subsequently scored by comparison to pre-set target values. Depth did not significantly differ between simulator and micro-computed tomography scan measurements (P = .139). Width differed significantly (P = .001), whereupon a significant regression equation was found (P < .0001) to calibrate the simulator. After calibration, no significant difference was found (P = .865). In total, 99 surgeons tested the simulator and more than agreed with the statements that the simulator is a good method for training MIMVS, and that the mitral valve and suture placement looked and felt realistic.<br />Conclusions: We successfully developed a high-fidelity MIMVS simulator for endoscopic and robotic approaches. The simulator provides a platform to train skills in an objective and reproducible manner. Future studies are needed to provide evidence for its application in training surgeons.<br /> (Copyright © 2018 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1097-685X
Volume :
157
Issue :
4
Database :
MEDLINE
Journal :
The Journal of thoracic and cardiovascular surgery
Publication Type :
Academic Journal
Accession number :
30385017
Full Text :
https://doi.org/10.1016/j.jtcvs.2018.09.014