Back to Search Start Over

Lead fracture of peripheral nerve stimulator for brachial plexopathy -a case report.

Authors :
Choi SC
Chae JS
Kim YJ
Chon JY
Moon HS
Source :
Korean journal of anesthesiology [Korean J Anesthesiol] 2018 Oct; Vol. 71 (5), pp. 407-410. Date of Electronic Publication: 2018 Jul 24.
Publication Year :
2018

Abstract

Peripheral nerve stimulation (PNS) is a useful treatment for chronic pain, but it can cause damage depending on its application site. Here, we describe the case of a 54-year-old man who underwent PNS for brachial plexopathy in 2015. One lead was implanted on the left medial cord to stimulate the medial antebrachial cutaneous nerve, and the other was implanted on the radial nerve to stimulate the posterior antebrachial cutaneous nerve. Both leads were inserted near the shoulder joint but did not cross it. Before PNS, the patient did not move his shoulder and elbow because of severe pain, but the treatment greatly alleviated this pain. Twenty months after the operation, both leads were fractured, and the severe pain returned. Repetitive motion near the joint was closely related to the lead fractures. In conclusion, large joints as the insertion sites of PNS leads should be avoided to prevent lead fractures.