Comparison of Enterprise Point-of-Care and Nova Biomedical Critical Care Xpress analyzers for determination of arterial pH, blood gas, and electrolyte values in canine and equine blood.

Authors :
Elmeshreghi TN
Grubb TL
Greene SA
Ragle CA
Wardrop JA
Source :
Veterinary clinical pathology [Vet Clin Pathol] 2018 Sep; Vol. 47 (3), pp. 415-424. Date of Electronic Publication: 2018 Jul 09.
Publication Year :
2018

Abstract

Background: Point-of-care analyzers can provide a rapid turnaround time for critical blood test results. Agreement between the Enterprise Point-of-Care (EPOC) and bench-top laboratory analyzers is important to determine the clinical reliability of the EPOC.<br />Objectives: The aim of the study was (1) to evaluate the precision (repeatability) of blood gas values measured by the EPOC and (2) to determine the level of agreement between the EPOC and Nova Critical Care Express (Nova CCX) for the assessment of arterial pH, blood gases, and electrolyte variables in canine and equine blood.<br />Methods: Arterial blood samples from dogs were analyzed on the EPOC and Nova CCX analyzers to determine precision and agreement of pH, PaCO <subscript>2</subscript> , PaO <subscript>2</subscript> , and HCT. The same analytes plus Na <superscript>+</superscript> , K <superscript>-</superscript> , and Cl <superscript>-</superscript> were analyzed for agreement using equine blood. Statistical analyses included assessment of precision using the coefficient of variation (CV%), and agreement using the Deming regression, Pearson correlation, and Bland-Altman plots.<br />Results: Both analyzers provided precise results of pH, PaCO <subscript>2</subscript> , PaO <subscript>2,</subscript> and HCT, meeting CV% quality requirement values. In both species, Deming regression results were acceptable and correlation values were above 0.93 for arterial pH and blood gases, but lower for sodium and chloride. Bland-Altman plots demonstrated varying degrees of bias, but good agreement between the 2 analyzers was seen when arterial blood gases and electrolytes were measured, except for PaCO <subscript>2</subscript> and Cl <superscript>-.</superscript> CONCLUSION: The EPOC analyzer provides consistent, reliable results for canine arterial blood gas values and for equine arterial blood gas and electrolyte values. Cl <superscript>-</superscript> results could be acceptable with the application of a correction factor, but the PaCO <subscript>2</subscript> results were more variable.<br /> (© 2018 American Society for Veterinary Clinical Pathology.)

Details

Language :
English
ISSN :
1939-165X
Volume :
47
Issue :
3
Database :
MEDLINE
Journal :
Veterinary clinical pathology
Publication Type :
Academic Journal
Accession number :
29989207
Full Text :
https://doi.org/10.1111/vcp.12635