Back to Search Start Over

Multiple Thromboembolic Events from a Left Atrial Appendage Occlusion Device.

Authors :
Prosperi-Porta G
Schnell G
Colbert J
Franko A
Wilton SB
Kuriachan VP
Source :
The Canadian journal of cardiology [Can J Cardiol] 2018 Mar; Vol. 34 (3), pp. 342.e13-342.e15. Date of Electronic Publication: 2017 Dec 27.
Publication Year :
2018

Abstract

Left atrial appendage occlusion devices are an alternative to oral anticoagulation in patients with nonvalvular atrial fibrillation who are at risk of ischemic stroke. Thromboprophylaxis after implantation is recommended, but the optimal regimen is unknown. We report a clinicopathologic case in which thrombus adherent to an incompletely endothelialized WATCHMAN device (Boston Scientific, Marlborough, MA) resulted in multiple thromboembolic events, contributing to a fatal outcome. This case illustrates uncertainties regarding the device's endothelialization process.<br /> (Copyright © 2017 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1916-7075
Volume :
34
Issue :
3
Database :
MEDLINE
Journal :
The Canadian journal of cardiology
Publication Type :
Academic Journal
Accession number :
29475536
Full Text :
https://doi.org/10.1016/j.cjca.2017.12.017